Step impedance spectroscopy: High speed GITT

Home :: Step impedance spectroscopy: High speed GITT
Fig. 1 GITT analysis of lead acid battery

Fig. 1 GITT analysis of lead acid battery

Electrochemical impedance spectroscopy (EIS) is typified by sequential application of sinusoidal perturbations over a range of fixed frequencies from mHz to MHz. The time bottleneck is the low frequency measurements. A more advanced EIS excitation signal consists of multiple of sine-waves. The fast Fourier transformations (FFT) response analysis requires non-overlapping second harmonics, requiring expensive, complex signal generators. The need for complex generators is obviated by use of square wave excitation. The analysis of a current square wave is the basis for the galvanostatic intermittent titration technique (GITT). The FFT analysis is supplanted by analysis of time domain data in terms of analyst selected circuit models. GITT takes advantage of measurable boundary values associated with the start and end of the GITT excitation signal. EIS lacks this capability. GITT directly provides physically relevant model parameters from the fit of the time-domain data, such as the change of voltage with respect to state-of-charge, solid state diffusion relaxation (after the end of pulse), and more.

Consider analysis of 2 year old Panasonic (LC-R064R5P) 6V, 4.5Ah/20HR, lead acid battery by a Powerstat-05 potentiostat/galvanostat. Figure 1 shows the voltage response (black line) to a current square wave (red line) applied to the battery. The amplitude (ip) was 0.5 A (red line) with a pulse width (tpw) 0.5 seconds. The precipitous drop in voltage (ΔV1) corresponds to the Randles cell series resistance.   This information is equivalent to that obtained with the high frequency intercept of a Nyquist plot . The ΔEeq is due to the change in the state-of-charge occurring during the pulse duration (tpw). Such information is not accessible through EIS. If tpw is short, E(t) is approximately linear with respect to the state-of-charge.  This simplifies fitting of the decay during tpw for double layer capacitance and charge transfer resistance.  In the case of a simple Randles cell, the tpw decay has an RC time constant related to double layer capacitance and charge transfer resistance.  When the pulse signal reverts to zero current during the  post pulse relaxation period (trp), there is no further change in the state-of-charge.  A battery equivalent circuit (Reference 3) is shown (Fig. 1). The remainder of this treatment shows how the differential equation for the circuit is solved and used in the NuVant EZware software to deliver parameters specific to the circuit model. Screen shots of EZware demonstrate how circuit elements are easily obtained using any NuVant potentiostat/galvanostat for battery testing.

It is noteworthy, that end users must select a reasonable value for the trp.  If trp is too short, ΔEeq will be overestimated. For the lead acid battery of this study, the relaxation time is on the order of a second.  For a deep current pulse with a lithium ion battery, the relaxation time could be greater than days.  Diffusion coefficients in liquid phase are on the order of 10-6 cm2s-1.  Solid state diffusion coefficients are 4 – 5 orders of magnitude smaller:  The relaxation time for Li to redistribute in solid state cathodes is substantial longer than the relaxation time required for macroscopic lead acid battery electrodes to relax.  Thus much can be learned from a simple but powerful SIS analysis.  The ΔV3 current transient provides information concerning diffusional processes in electrochemical devices in addition to double layer capacitance.  NuVant provides additional equivalent circuit solutions addressing corrosion, fuel cells, batteries and more in the EZware software. SIS is an affordable and powerful alternative to EIS for equivalent circuit analysis.The data for the entire analysis below was acquired in less than a second.  Our instruments can be multiplexed such that thousands of batteries can be tested by a single instrument in one day.

 

Glossary Abbreviations for GITT
t0   Beginning of the pretreatment period
tp,i   Pulse initiation time
tp,f   Pulse final time
tr,f   End of relaxation time
tpw = tpf  – tpi   Pulse width
trp = trf  – tpf   Relaxation period
E(t) =  Eeq1   For ( tp,i – tinterv ≤ t ≤ tp,i )
Eeq2   ~E(tr,f  )
Eeop = E(tp,f)   Potential at end of pulse
τ   RC time constant specific to differential equation
ip   Pulse amplitude
tinterv   Sampling interval time

 

Battery Equivalent Circuit
battery circuit
Figure 2.  Equivalent circuit for lead acid battery analysis

 

The differential equation applicable to the circuit is below:
Battery eq1
battery eq2
battery eq4
battery eq3
Current pulse solution to the differential equation:
battery eq5
When I=ip (tpi ≤ t ≤ tpf)
battery eq6
When I= O (tpf ≤ t ≤ trf)
Where battery eq7  and tpw=tpf – tpi

 

Zcp(t) and E(t) vary linearly with t when tpw is short (i.e., E(t) contributes a linear component to ΔV2). EZware uses Eeq1 and Eeq2 to calculate ΔEeq.     ΔV1 = f(Rs, ip)   ΔV2 = f(Rct, E(t), Cdl)   ΔV3 = f(Rct , Cdl)

EZware-2013 Software

Recipe EditorSIS recipe editor Data CollectionSIS data window Data AnalysisSIS data analysis battery

GITT  or Step Impedance Spectroscopy References

  1. Barsoukov, E.; Ryu, S. H.; Lee, H., A novel impedance spectrometer based on carrier function Laplace-transform of the response to arbitrary excitation. Journal of Electroanalytical Chemistry 2002, 536 (1–2), 109-122.
  2. Ferrese, A.; Albertus, P.; Christensen, J.; Newman, J., Lithium Redistribution in Lithium-Metal Batteries. Journal of The Electrochemical Society 2012, 159 (10), A1615-A1623.
  3. Gagneur, L.; Driemeyer-Franco, A. L.; Forgez, C.; Friedrich, G., Modeling of the diffusion phenomenon in a lithium-ion cell using frequency or time domain identification. Microelectronics Reliability 2013, 53 (6), 784-796.
  4. Kuhn, E.; Forgez, C.; Lagonotte, P.; Friedrich, G., Modelling Ni-mH battery using Cauer and Foster structures. Journal of Power Sources 2006, 158 (2), 1490-1497.
  5. Milocco, R. H.; Castro, B. E., State of charge estimation in Ni-MH rechargeable batteries. Journal of Power Sources 2009, 194 (1), 558-567.
  6. Neumann-Spallart, M.; Etman, M., Calculation of impedance spectra by Laplace transformation of voltage transients generated by current-step excitation. Journal of Electroanalytical Chemistry 1994, 372 (1–2), 33-37.
  7. Onda, K.; Nakayama, M.; Fukuda, K.; Wakahara, K.; Araki, T., Cell Impedance Measurement by Laplace Transformation of Charge or Discharge Current–Voltage. Journal of The Electrochemical Society 2006, 153 (6), A1012-A1018.
  8. Pilla, A. A., A Transient Impedance Technique for the Study of Electrode Kinetics: Application to Potentiostatic Methods. Journal of The Electrochemical Society 1970, 117 (4), 467-477.
  9. Pukrushpan, J. T.; Peng, H.; Stefanopoulou, A. G., Control-oriented modeling and analysis for automotive fuel cell systems. J Dyn Syst-T Asme 2004, 126 (1), 14-25.
  10. Remmlinger, J.; Buchholz, M.; Meiler, M.; Bernreuter, P.; Dietmayer, K., State-of-health monitoring of lithium-ion batteries in electric vehicles by on-board internal resistance estimation. Journal of Power Sources 2011, 196 (12), 5357-5363.
  11. Savoye, F.; Venet, P.; Millet, M.; Groot, J., Impact of Periodic Current Pulses on Li-Ion Battery Performance. Industrial Electronics, IEEE Transactions on 2012, 59 (9), 3481-3488.
  12. Thele, M.; Bohlen, O.; Sauer, D. U.; Karden, E., Development of a voltage-behavior model for NiMH batteries using an impedance-based modeling concept. Journal of Power Sources 2008, 175 (1), 635-643.
  13. Vilar, E.; Dougal, R. A., Study of pulsed-current loading of direct methanol fuel cells using a new time-domain model based on bi-functional methanol oxidation kinetics. Journal of Power Sources 2007, 169 (2), 276-287.
  14. Yan, Q. G.; Toghiani, H.; Causey, H., Steady state and dynamic performance of proton exchange membrane fuel cells (PEMFCs) under various operating conditions and load changes. Journal of Power Sources 2006, 161 (1), 492-502.
  15. L. Redey, J. Prakash, D.R. Vissers and K.M. Myles, Development of High-Performance Na/NiCl2 Cell”, Proceedings of the 35th International Power Sources Symposium,” Cherry Hill, NJ, June 22, pp. 343, 1992. 16. “Electrochemical Behavior of Nonporous Ni/NiCl2 Electrode in Chloroaluminate Melts,” Jai Prakash, Laszlo Redey, and Donald Vissers, J. Electrochem. Soc., 147, 502 (2000).